Analysis of Exposure Uniformity on Cells in Vitro with 900MHz TEM Chamber

Full Text (PDF, 404KB), PP.86-93

Views: 0 Downloads: 0

Author(s)

J. N. ZHANG 1 J. X. ZHAO 1 H. M. LU 1 W. WANG 1 Z. REN 1

1. School of Electronic Engineering, Xidian University, No. 2 South Taibai Road, Xi’an 710071, China

* Corresponding author.

DOI: https://doi.org/10.5815/ijem.2011.05.12

Received: 15 Jun. 2011 / Revised: 1 Aug. 2011 / Accepted: 5 Sep. 2011 / Published: 5 Oct. 2011

Index Terms

TEM Chamber, Exposure Uniformity, Specific Absorption Rate (SAR)

Abstract

The TEM chamber is a proposed exposure device for experiments detecting cell responses to electromagnetic irradiation at the mobile frequency of 900 MHz. In this paper, we extend the traditional propagating-wave exposure to the standing-wave exposure. Four standing waves with different properties are produced respectively by two types of passive loads and two types of active loads. Systematically, the maximum E field and maximum H field of the standing waves are used to expose Petri dishes of three sizes. The standard deviation (SD) of the specific absorption rate (SAR) is calculated for the cell layer and the cell suspension. We compare the SD data to determine the best arrangement with the highest exposure uniformity while considering the exposure intensity. The investigation uses the numerical simulation of the finite-difference time-domain (FDTD) method with the result accuracy supported by comparison to theoretical data on the field distribution in the TEM chamber.

Cite This Paper

J. N. ZHANG, J. X. ZHAO, H. M. LU, W. WANG, Z. REN,"Analysis of Exposure Uniformity on Cells in Vitro with 900MHz TEM Chamber", IJEM, vol.1, no.5, pp.86-93, 2011. DOI: 10.5815/ijem.2011.05.12 

Reference

[1]F. Schönborn, K. Poković, M. Burkhardt, and N. Kuster, “Basis for optimization of in vitro exposure apparatus for health hazard evaluations of mobile communications,” Bioelectromagnetics, vol. 22, pp. 547-559.

[2]G. Prisco, G. d’Ambrosio, M. Calabrese, R. Massa, and J. Juutilainen, “SAR and efficiency evaluation of a 900 MHz waveguide chamber for cell exposure,” Bioelectromagnetics, vol. 29, pp. 429-438.

[3]M. Popović, S. Hagness, and A. Taflove, “Finite-difference time-domain analysis of a complete transverse electromagnetic cell loaded with liquid biological media in culture dishes,” IEEE Trans. Bio-med. Eng., vol. 45, pp. 1067-1076.

[4]ICNIRP, “Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields (up to 300 GHz),” Health Physics, vol. 74, pp. 494-522.

[5]J. Schuderer and N. Kuster, “Effect of the meniscus at the solid/liquid interface on the SAR distribution in Petri dishes and flasks,” Bioelectromagnetics, vol. 24, pp. 103-108. 

[6]J. Schuderer, D. Spät, T. Samaras, W. Oesch, and N. Kuster, “In vitro exposure systems for RF exposures at 900 MHz,” IEEE Trans. Microwave Theory Tech., vol. 52, pp. 2067-2075.